Thermoelectric Cooling Solutions

site map eng | ru

Electronic Devices

Print version

Gas Analyzers Specifications

SPECIFICATION

ParameterUnitsModel
CO2 CnHm CH4
a b c
Type NDIR dual channel sensor double pass optical scheme
Detector Integrated Pair of Lead Selenide Detector and Two Single Solid State Light Emitters. The Optopair with two-stage TE cooler.
Light Source
Sampling Mode Aspiration
Parameters
Concentration Range2) ppm/%vol 0...1000ppm 0...5%vol 0...20%vol 0...1000ppm 0...5%vol
Noise level3,4) ppm/% <10 ppm <0.5% 0.5% <10 ppm <1%
Accuracy4) ppm/% 20 ppm 1% 1% 20 ppm 3%
Zero Drift4) %0.1
Sampling Cell Volume5) cm3 1.0
Timing
Output repeating Rate 1) Hz 0.01...10
Averaging Time Constant1) s 0.1...60
Alarms
Light - Two-color LED
Sound1) dB >85
Supply requirements
Supply Voltage V DC +12
Current, max mA 300
Output Interfaces
Digital - RS232C
Current, max mV 0...4095
Operation Conditions
Temperature Range °C -10...+50
Relative Humidity % 5 ... 95
Mechanical
Dimensions mm3 59 x 71 x 79 (LCD)
46x71x79 (Mini)
42x73x92 (OEM with covers)
Weight g 300 (LCD)
240 (Mini)
180 (OEM)

1) Software Adjustable

2) Optional ranges up to 100% vol. are available.

3) At Averaging Time Constant =0.2 sec.

4) Value in % means relative units DX/X