DX6220 SERIES OF GAS ANALYZERS


Up ]
[Download Manual]

SPECIFICATIONS

Parameter Units Model
CO2 CnHm CH4
a b c
Type  

NDIR dual channel sensor double pass optical scheme

Detector   Integrated Pair of Lead Selenide Detector  and Two Single Solid State Light Emitters. The Optopair with two-stage TE cooler. 
Light Source  
Sampling Mode   Aspiration
Parameters
Concentration Range2) ppm/%vol 0...1000ppm 0...5%vol 0...20%vol 0...1000ppm 0...5%vol
Noise level 3,4) ppm/% <10 ppm <0.5% 0.5% <10 ppm <1%
Accuracy4) ppm/% 20 ppm 1% 1% 20 ppm 3%

Zero Drift4)

% 0.1

Sampling Cell Volume 5)

cm3 1.0

Timing

Output repeating Rate 1) Hz 0.01...10

Averaging Time Constant1)

s 0.1...60 

Alarms

Light  - Two-color LED
Sound dB >85 
Supply requirements
Supply Voltage V DC +12
Current, max mA 300
Output Interfaces
Digital  - RS232C
Analog mV 0...4095
Operation Conditions
Temperature Range oC -10...+50
Relative Humidity % 5 ... 95
Mechanical
Dimensions mm3 59 x 71 x 79 (LCD)
46x71x79 (Mini)
42x73x92 (OEM with covers)
Weight g 300 (LCD)
240 (Mini)
180 (OEM)

1) Software Adjustable
2) Optional ranges up to 100% vol. are available.
3) At Averaging Time Constant =0.2 sec.
4) Value in % means relative units
DX/X


Information furnished by RMT Ltd is believed to be reliable, however, no responsibility is assumed for possible inaccuracies or omission. Specifications are subject to change without notice.

 
 